H 2 O 3 - ISH(PROIII) thermoelectric (Peltier) Slide Hybridizer allows you to program incubation temperature and time duration in three segments. The temperature can be set to between 0 ºC~100 ℃, and the time up to 24 hours for each segment. H 2 O 3 -ISH(PROIII) has a modular design of blocks to accommodate 5 pieces of glass slide.
Contents of Package
- One thermoelectric incubator
- One 12 V power supply adaptor
- One sample block cover
- One Flat Block
- One Plastic Groove
- One User Manual
Specifications
- Temperature Range: 0~100ºC
- Time Range: 1 min to 23 hr 59 min each segment,
- Programmable segments: 3
- Temperature precision:±0.2 ℃
- Display precision:0.1 ℃
- Sample block temperature evenness:±0.5 ℃
- Operation ambient temperature: 8~30ºC
- Dimensions: 185×185×90 mm (7.25×7.25×3.5 in)
- Net Weight (with blocks): ~2.5 kg
Operation Instruction of H 2 O 3 -ISH((PROIII)
Setting up Place the device on your lab bench. There should be at least two inches (5 cm) of clearance around the device to allow proper ventilation. Make sure that the space between the bench and the unit is also clear.
Place aluminum block on Cold/Hot Plate (Fig. 1). Make certain that the sample blocks are in close contact with the plate. Plug in the provided 12 V DC power supply to a wall outlet, and connect the DC plug to the DC input terminal on the back of the unit(Fig. 2). Turn on the power switch (Fig. 2). Upon turning on the unit, Timer Display shows “OFF”
and Temperature Display shows the real temperature of Cold/Hot Plate. Now the unit is in
OFF mode (in which the heating/chilling core is inactive). Your are ready to program the
device.
Programming The temperature profile of H 2 O 3 -ISH(PROIII )has 3 segments to program. Press “SET” key to enter the programming mode to set the temperature and time duration of each segment. For example, press “SET” once, Timer Display starts flashing and the first (leftmost) Segment LED turns on; they indicate that the time of the first segment is selected to program. Now, press up ▲ or down▼ key to increase or decrease the time. Press “SET” again, Temperature Display starts flashing; it means that the temperature of the first segment is selected to program. Press “SET” key once again, Timer Display flashes and the second (middle) Segment LED turns on; now the time of the second segment is ready to program. So on, continue to press “SET” key to advance parameter and segment selection. Once a parameter is selected, press up ▲ or down▼ key to adjust the value of that parameter.
Running Press “RUN” button to start the program. When the temperature reaches within ±1.0 ℃ from the set point, the buzzer beeps once, the time countdown begins and Timer Display colon (:) starts flashing. Upon completion of all three segments, the buzzer beeps three times, and the device returns to OFF mode.